Free Chlorine
Description
Versions
Downloads
Description
Electrochemical sensor for measurement of chlorine concentrations from inorganic chlorine products in water.
The chlorine measuring probe is an electrochemical sensor for measuring the concentration of chlorine in water. The measuring cell captures free chlorine from inorganic chlorine products (hypochlorite, chlorine gas, etc.). The measuring method has a reduced pH dependency, so that pH fluctuations only have a limited impact on the measurement signal. pH value increases only lead to an approximately 10 % reduction of the measuring signal per pH unit. The Sensor has a plug and is not delivered with a fixed cable.
Versions
Downloads
Benefits
- Stable signals even with fluctuating pH values
- Abrasive particles are tolerated
- Surfactants are partially tolerated
Areas of application
- Swimming pools, drinking water, seawater
Parameters
- Free Chlorine
Request now
Do you need individual advice or are you interested in the chlorine sensor?
Technical Data
Technical Data
Technical Data
Measurement principle | Amperometry | |
Measuring range | 0…2 mg/L, 0…20 mg/L | |
Measurement accuracy | Measurement range 2 mg/L: at 0.4 mg/L; 1.6 mg/L < 1 % | |
Measurement range 20 mg/L: at 4 mg/L < 1 %; at 16 mg/L W37< 3 % | ||
Temperature compensation | automatic through integrated temperature sensor | |
Response time | T90: ~ 2 min | |
Housing material | Micro-porous hydrophylic membrane, PVC-U, Stainless steel 1.4571 | |
Dimensions (L x Ø) | 205 mm x 25 mm | ~ 8˝ x 1˝ |
Weight | 0,1 kg | ~ 0.2 lbs |
Interface | RS-485, Modbus RTU | |
Power supply | 12…24 VDC (± 10 %) | |
Connector | 8-pin M12-plug | |
Power consumption | 0.6 W | |
Maintenance effort | ≤ 0.5 h/month (typical) | |
Calibration/maintenance interval | Membrane cap change: every 12 months | |
Electrolyte change: every 12 months | ||
System compatibility | Modbus RTU | |
Warranty | 1 Year (EU & US: 2 years) on electronics; All wearing parts are not included in the warranty | |
Max. pressure | 3 bar | ~ 43.5 psig |
Protection type | IP68 | NEMA 6P |
Sample temperature | 0…+45 °C | ~ +32 °F… +113 °F |
Ambient temperature | 0…+55 °C | ~ +32 °F… +131 °F |
Storage temperature | +5…+35 °C | ~ +41 °F… +95 °F |