LISA color
State-of-the-art VIS absorption photometer for highly precise SAC and colour measurements. The integrated LED-Technology ensures a long product lifetime.
Colorimetry – LISA enables reliable low-cost color measurements. LISA color uses two different LEDs for long-term stable measurements of SAC or colors at different wavelengths. The second channel is used for turbidity/background correction. The cutting-edge device platform, used in all other TriOS photometers, enables optical path lengths of 50, 100, 150 and 250 mm, so that almost any application can be easily implemented. LISA color also enables applications in aggressive media (e.g. high chloride concentrations) thanks to the optional titanium housing. Equipped with our G2 interface with web browser configuration, internal data logger, flexible protocols and data outputs, LISA color includes features that are much more advanced than those of comparable devices currently available on the market. The unified platform of all TriOS photometers also facilitates a standardized spare parts and consumables system, which allows the use of a wide range of accessories for our devices. Furthermore the cutting-edge G2 interface enables quick integration into third-party systems. |
Light Source | 2 LEDs | |
Detector | Photo diode | |
Measurement principle | Attenuation, Transmission | |
Optical path | 50 mm, 100 mm, 150 mm, 250 mm | |
Parameter | SAC436 or Colouring (based on DIN EN ISO 7887 (410 nm, 436 nm, 525 nm or 620 nm)) or Pt-Co color number (APHA/Hazen) (390 nm or 455 nm) or Cr-Co color number (390 nm or 413 nm) | |
Measurement accuracy | 0.5 % | |
Turbidity compensation | at 740 nm | |
Data logger | ~ 2 MB | |
T100 response time | min. 4 s | |
Measurement interval | min. 2 s | |
Housing material | Stainless steel (1.4571/1.4404) or titanium (3.7035) | |
Dimensions (L x Ø) | 340 mm x 48 mm (with 50 mm path) | ~ 13.4” x 1.9” (with 50 mm path) |
Weight stainless steel | ~ 2.4 kg (with 50 mm path) | ~ 5.3 lbs (with 50 mm path) |
Weight titanium | ~ 1.3 kg (with 50 mm path) | ~ 2.9 lbs (with 50 mm path) |
Interface digital | Ethernet (TCP/IP), RS-232 or RS-485 (Modbus RTU) | |
Interface analog | Ethernet (TCP/IP), 4...20 mA | |
Power consumption | ≤ 1 W | |
Power supply | 12...24 VDC (± 10 %) | |
Maintenance effort | ≤ 0.5 h/month (typical) | |
Calibration/maintenance interval | 24 months | |
System compatibility | Modbus RTU or Analog Out (4...20 mA) | |
Warranty | 1 year (EU: 2 years) | US: 2 years |
Max. pressure with SubConn | 30 bar | ~ 435 psig |
Max. pressure with fixed | 3 bar | ~ 43.5 psig |
Max. pressure in FlowCell | 1 bar, 2...4 L/min | ~ 14.5 psig, 0.5 to 1.0 gpm |
Protection type | IP68 | NEMA 6P |
Probentemperatur | +2...+40 °C | ~ +36 ˚F to +104 ˚F |
Umgebungstemperatur | +2...+40 °C | ~ +36 ˚F to +104 ˚F |
Lagertemperatur | -20...+80 °C | ~ -4 ˚F to +176 ˚F |
Anströmgeschwindigkeit | 0,1...10 m/s | ~ 0.33 fps to 33 fps |
Parameter Variations | According to the standard | Unit | Factor | Path Length [mm] | ||||
10 | 50 | 100 | 150 | 250 | ||||
SAC 436 nm | DIN EN ISO 7887:2011_method B | 1/m | - | 0.5...150 | 0.1...30 | 0.05...15 | 0.03...10 | 0.02...6 |
SAC 525 nm | DIN EN ISO 7887:2011_method B | 1/m | - | 0.5...150 | 0.1...30 | 0.05...15 | 0.03...10 | 0.02...6 |
SAC 620 nm | DIN EN ISO 7887:2011_method B | 1/m | - | 0.5...150 | 0.1...30 | 0.05...15 | 0.03...10 | 0.02...6 |
True Color 410 nm | DIN EN ISO 7887:2011_method C | mg/L Pt | 18.52 | 10...2800 | 2...560 | 1.0...280 | 0.6...185 | 0.4...110 |
Pt-Co-Color 390 | DIN EN ISO 6271-2016:05 | mg/L Pt | 7.4 | 4...1100 | 0.8...220 | 0.4...110 | 0.3...75 | 0.2...45 |
Pt-Co-Color 455 | DIN EN ISO 6271-2016:05 | mg/L Pt | 36.4 | 20...5500 | 4...1100 | 2.0...550 | 1.5...360 | 0.8...220 |
Cr-Co-Color 380 | - | ° (color degree) | 9.7 | 5...1500 | 1...300 | 0.5...150 | 0.3...100 | 0.2...60 |
Cr-Co-Color 413 | GOST 3351:1974 | ° (color degree) | 34.1 | 20...5500 | 4...1100 | 2.0...550 | 1.5...360 | 0.8...220 |
Turbidity 740 nm | - | FAU** | 6.0 / 0.01242 | 3...330 | 0.6...60 | 0.3...30 | 0.2...20 | 0.12...12 |
under laboratory conditions **Formazin Attenuation Unit | ||||||||