Technical specifications
Measurement technology | Light source | 2 LEDs (254 nm, 530 nm) |
---|---|---|
Detector | Photodiode | |
Measuring principle | Attenuation, transmission | |
Optical path | 0.3 mm, 1 mm, 2 mm, 5 mm, 10 mm, 20 mm, 50 mm, 100 mm |
Parameters | SAC254, CODeq, BOD eq, TOCeq, UVT, Turb530 |
---|---|
Measuring range | See table of parameters |
Repeatability | 0.2 % |
Turbidity compensation | at 530 nm |
Data logger | ~ 2 MB |
Response time T100 | 4 s |
Measuring interval | ≥ 2 s |
Housing material | Stainless steel (1.4571/1.4404) or titanium (3.7035) | ||
---|---|---|---|
Dimensions (L x Ø) | 300 mm x 48 mm (with 10 mm path) | ~ 11.8'' x 1.9'' (with 10 mm path) | |
Weight | VA | ~ 2.3 kg (with 10 mm path) | ~ 5.1 lbs (with 10 mm path) |
Ti | ~ 2.1 kg (with 10 mm path) | ~ 4.6 lbs (with 10 mm path) |
Interface | digital | Ethernet (TCP/IP) |
---|---|---|
RS-232 or RS-485 (Modbus RTU) | ||
analog | 4 .. 20 mV, max. load: 500 Ohm | |
Power consumption | ≤ 1 W | |
Power supply | 12 – 24 VDC (± 10 %) |
Maintenance effort | ≤ 0.5 h/month (typical) |
---|---|
Calibration/maintenance interval | 24 months |
System compatibility | Modbus RTU or: Analog output (4 .. 20 mA) |
Warranty | 1 year (EU & USA: 2 years) |
INSTALLATION
Max. | with SubConn | 30 bar | ~ 435 psi |
---|---|---|---|
with fixed cable | 3 bar | ~ 43.5 psi | |
in flow unit | 1 bar, 2...4 L/min | ~ 14.5 psi, 2 to 4 L/min | |
Degree of protection | IP68 | NEMA 6P |
Sample temperature | +2...+40 °C | ~ +36 to +104 °F |
---|---|---|
Ambient temperature | +2...+40 °C | ~ +36 to +104 °F |
Storage temperature | -20...+80 °C | ~ -4 to +176 °F |
Inflow velocity | 0.1...10 m/s | ~ 0.33 to 33 fps |