Technical specifications
Measurement technology | Light source | Xenon flash lamp |
|---|---|---|
Detector | High-end miniature spectrometer | |
256 channels | ||
200 to 360 nm | ||
0.8 nm/pixel | ||
Optical path | OPUS | 0.3 mm, 1 mm, 2 mm, 5 mm, 10 mm, 20 mm, 50 mm |
OPUS aero | 0.3 mm, 1 mm, 2 mm, 5 mm |
Parameters | OPUS | See paragraph “Measurement ranges and detection limits OPUS” |
|---|---|---|
OPUS aero | Nitrate NO3-N or nitrate NO3-N + nitrite NO2-N | |
Measurement range | OPUS | See paragraph “OPUS measuring ranges dep. on the path length” |
OPUS aero | See paragraph “OPUS aero measuring ranges” | |
Measurement accuracy | OPUS | See paragraph “Measurement ranges and detection limits OPUS” |
OPUS aero | ± (5 % + 0,1) | |
Turbidity compensation | Yes | |
Data logger | ~ 2 GB | |
Response time T100 | 2 min | |
Measurement interval | ≥ 1 min | |
Housing material | Stainless steel (1.4571/1.4404) or titanium (3.7035) | ||
|---|---|---|---|
Dimensions (L x Ø) | 470 mm x 48 mm* | ~ 18.5" x 1.9" | |
Weight | VA | ~ 3 kg* | ~ 6.6 lbs |
Ti | ~ 2 kg* | ~ 4.4 lbs | |
Interface | digital | Ethernet (TCP/IP) |
|---|---|---|
RS-232 or RS-485 (Modbus RTU) | ||
Power consumption | ≤ 8 W | |
Power supply | 12-24 VDC (± 10 %) | |
Maintenance effort | ≤ 0.5 h/month typical |
|---|---|
Calibration/ | 24 months |
System compatibility | Modbus RTU |
Warranty period | 1 year (EU: 2 years) |
Sample temperature | 0...+55 °C* | ~ +32 to +131 °F* |
|---|---|---|
Ambient temperature | 0...+55 °C* | ~ +32 to +131 °F* |
Storage temperature | -20...+80 °C | ~ -4 to +176 °F |
Approach velocity | 0.1...10 m/s | ~ 0.33 t 33 fps |
Max. pressure | with Subconn* | 30 bar | ~ 435 psi |
|---|---|---|---|
with fixed cable | 3 bar | ~ 43.5 psi | |
in flow unit | 1 bar, 2...4 L/min | ~ 14.5 psi at 0.5 to 1.0 gpm | |
Degree of protection | IP68 | ||