Technical specifications

Measurement technology

Light source

2 LEDs

Detector

photodiode

Measuring principle

Attenuation, Transmission

Optical path

1 mm, 2 mm, 5 mm, 10 mm, 50 mm, 100 mm, 150 mm, 250 mm

Parameters

SAK436, SAK525, SAK620

Color (based on DIN EN ISO 7887 (410 nm, 436 nm, 525 nm, 620 nm)

Pt-Co color number (APHA/Hazen) (390 nm or 455 nm)

Cr-Co color number (380 nm or 413 nm)

Measuring range

See parameter list for measuring ranges below

Measuring accuracy

0,5 %

Turbidity compensation

Yes, 740 nm

Data logger

~ 2 MB

Response time T100

4 s

Measuring interval

≥ 2 s

Table: *for 50-mm path

Housing material

Stainless steel (1.4571/1.4404) or titanium (3.7035)

Dimensions (L x Ø)

340 mm x 48 mm*

~ 13.4" x 1.9" *

Weight

VA

~ 2.4 kg*

~ 5.3 lbs*

Ti

~ 1.3 kg*

~ 2.9 lbs*

Interface

digital

Ethernet (TCP/IP)

RS-232 or RS-485 (Modbus RTU)

analog

Ethernet (TCP/IP)

4 .. 20 mA, max. load: 500 Ohm

Power consumption

≤ 1 W

Power supply

12 - 24 VDC (± 10 %)

Maintenance effort

≤ 0.5 h/month typical

Calibration/ maintenance interval

24 months

System compatibility

Modbus RTU

Analog Out (4 .. 20 mA)

Warranty

1 year (EU: 2 years)

Max. pressure

with SubConn

30 bar

~ 435 psi

with fixed cable

3 bar

~ 43.5 psi

in flow unit

1 bar, 2...4 L/min

~ 14.5 psi, 0.5 to 1 gpm

Degree of protection

IP68

NEMA 6P

Sample temperature

+2...+40 °C

~ +36 to +104 °F

Ambient temperature

+2...+40 °C

~ +36 to +104 °F

Storage temperature

-20...+80 °C

~ -4 to +176 °F

Inflow velocity

0.1...10 m/s

~ 0.33 fps to 33 fps