Technical specifications

TpH

Measurement technology

pH electrode

Measuring principle

Potentiometry

Parameters

pH value, temperature

Measuring range

pH

0...14 pH

Temperature

0...+65 °C

pH resolution

pH

0.01 pH

Temperature

0,1 °C

Measuring accuracy

pH

± 0.06 pH

temperature

± 0,5 °C

Intrinsic deviation

pH1

± 0.05 pH

pH7

± 0.05 pH

pH13

± 0.35 pH

Linearity measurement error

± 0.1 pH

Repeatability

pH1

± 0.1 pH

pH7

± 0.05 pH

pH13

± 0.1 pH

Output signal fluctuation

pH7

± 0.025 pH

pH4

± 0.05 pH

Warm-up time

< 5 min

Drift

Short-term drift 24 h

≤ 0.03 pH

Long-term drift 1 week

≤ 0.05 pH

10 % time and 90 % time

T10 increasing

< 2 s

T10 decreasing

< 2 s

T90 increasing

≤ 5 s

T90 waste decreasing

≤ 5 s

Temperature compensation

Pt1000

Measuring interval

2 s

Housing material

PPS / PET / NBR

Dimensions (L x Ø)

~ 180 x 27 mm

~ 7.1˝ x 1.1˝

Weight

110 g

~ 0.2 lbs

Interface

RS-485, Modbus RTU

Power consumption

0,2 W

Power supply

12 - 24 VDC (± 10 %)

Connection

8-pin M12 plug

Sensor cable

2 m and 10 m

Required supervision

Typically, ≤ 0.5 h/month

Calibration / maintenance interval

Typically, 4 weeks

System compatibility

Modbus RTU

Warranty

1 year (EU & USA 2 years) on electronics; wearing parts are excluded from the warranty

Max. pressure

with fixed cable

3 bar

~ 43.5 psi

in FlowCell

1 bar, 2...4 L/min

~ 14.5 psi, 0.5 to 1 gpm

Degree of protection

IP68

NEMA 6P

Sample temperature

+2...+40 °C

~ +36 to +104 °F

Ambient temperature

+5...+55 °C

~ +23 to +131 °F

Storage temperature

0...+80 °C

~ +32 to +176 °F

Inflow velocity

0...3 m/second

~ 0...10 fps