Technical specifications
Application | Long-term monitoring and limit value control of processes; water treatment |
---|---|
Measurement technology | Redox electrode |
Measuring principle | potentiometry |
Parameters | Oxidation-reduction potential |
Applied standard | DIN EN ISO 27888:1993 |
Measuring range | ± 1000 mV |
---|---|
Resolution | 0.01 mV |
Calibration | Calibration of the sensor is not necessary. |
Smallest measuring interval | ≥ 2 sec |
Temperature compensation | No |
Turbidity compensation | No |
Data logger | No |
Interface | digital: | RS-485, Modbus RTU |
---|---|---|
analog: | - | |
Power supply | 12-24 VDC (± 10 %) | |
Optical display | Status LED | |
Connection | 8-pin M12 plug | |
Sensor cable | 0.5 m, 2 m and 10 m |
Housing material | PPS, PET, NBR | |
---|---|---|
Dimensions (L x Ø) | ~180 mm x 27 mm | ~ 7.1 '' x 1.06 '' |
Weight | 110 g | 0.243 lbs |
Operating conditions | Temperature: | 0° C...+100 °C | 32...212 °F |
---|---|---|---|
Fixed cable: | 3 bar | 43.5 psi | |
Flow unit: | 1 bar at 2...4 L/min | 14.5 psi at 2...4 L/min | |
Protection class | IP68 | NEMA 6P |
Maintenance effort | < 0.5 h/month typical |
---|---|
System compatibility | TriBox3, TriBox mini, Modbus RTU |
Warranty | 1 year (EU & USA 2 years) on electronics; wearing parts are excluded from the warranty |
Accessories (optional) | Cables: | Extension cable 2 m, 10 m |
---|---|---|
Controller: | TriBox3, TriBox mini | |
Fittings: | FC eCHEM flow cell |