Technical specifications

Power supply

Voltage supply

100 - 240 VAC, 50 - 60 Hz

Power consumption

Max. 50 W

Interfaces

Digital

Ethernet

Analog

6 outputs: 4 .. 20 mA

Load

max. 500 Ω

Protocol

Modbus TCP/IP

Parameters

  • PAH
  • pH (BS EN 60746-2:2003)
  • Turbidity (DIN EN ISO 7027:2016)
  • Temperature (from TpH-D)
  • Flow rate (internal)
  • PAH turbidity corrected

Mechanic system

Size (width x height x depth)

600 x 800 x 337 mm

~ 23.6" x 31.5" x 13.3"

Weight

43 kg (without sensors)
45.5 kg (with sensors)

~ 95 lbs
~ 100 lbs

Environment

Sample temperature

+2 °C...+40 °C

~ +36 to +104 °F

Ambient temperature

0 °C...+45 °C

~ +32 to +113 °F

Storage temperature

-20 °C...+80 °C

~ -4 to +176 °F

Relative humidity

0...95 % (non-condensing)

pH value

> pH4

Degree of protection

IP56

Inlet

Max. pressure

Inlet pressure

1 to max. 25 bar

~ 14.5 psi to max 363 psi

Internal pressure

max. 3 bar

max. ~ 43.5 psi

Flow volume

2 - 5 L/min

Internal volume

Approx. 1 L