Technical specifications

Application

Determination of pH value and conductivity in drinking water

Measurement technology - pH

pH electrode

Measurement technology - EC

Conductivity

Measuring principle - pH

Potentiometry

Measuring principle - EC

Conductivity with two graphite electrodes

Parameters

pH value, conductivity, temperature

Applied standard

DIN EN ISO 27888:1993

pH

EC

Temperature

Measuring range

0...14 pH

0.00...5000 μS/cm

0...65 °C

32...149 °F

Measuring accuracy

± 0.06 pH

± 40 μS/cm at 1000 μS/cm, ± 200 μS/cm at 5000 μS/cm

± 0,5 °C

± 32.9 °F

Resolution

0.01 pH

<100 = 0.01 μS/cm; <1000 = 0.1 μS/cm; >1000 = 1 μS/cm

0,1 °C

32.2 °F

Repeatability precision

pH 1: 0.001; pH 7: 0.0006; pH13: 0.001

± 2 μS/cm at 1000 μS/cm; ± 7 μS/cm at 4000 μS/cm

-

Detection limit

Not
applicable

3 μS/cm

-

Response time T90 / T100

T90 rising 15 s ; T90 falling 20 s

Warm-up time

< 5 min

Stability / drift

Short-term drift 24 h: < 0.03 pH

Long-term drift 1 week: < 0.05 pH

Temperature compensation

Pt1000

Turbidity compensation

No

Data logger

Internal 8 GB memory

Response time

10 s

Shortest measurement interval

5 s

Cross sensitivities

None known

Display

3.5 inch capacitive color touch display, 320x480 pixels

Interface - digital

RS-485 (Modbus RTU), Ethernet (Modbus TCP)

Interface - analog

-

Power supply

12-24 VDC (± 10 %)

Power consumption

Typical 2 W; standby: 1.5 W

Protection class

CLASS III

Overvoltage category

I

Connection

M12 hybrid industrial connector, 8-pin

Materials

See dimension drawings (D10) in the operation instructions of the device

Dimensions (W/H/D)

160 / 280 / 108 mm

6.3 '' / 11 '' / 4.25 ''

Weight

Approx. 3.8 kg

Approx. 8.4 lbs

Operating conditions

Operating temperature

Sample (insitu) 2...40 °C

35.6...104 °F

Min. internal pressure

0.2 bar

2.9 psi

Max. internal pressure

1 bar

14.5 psi

Inflow velocity

Min. 10 L/h

Transport conditions

0...80 °C

32...176 °F

Storage conditions

0...80 °C

32...176 °F

Protection class

IP30

Maintenance effort

≤ 0.5 h / month typical

Calibration/
maintenance interval

pH: 4 weeks typical, EC: 6 months typical

Regular cleaning, depending on the water quality

System compatibility

TW Master, Modbus RTU, Modbus TCP

Warranty period

1 year (EU & USA: 2 years)