Technical specifications
Measurement technology | Light source | Xenon flash lamp |
|---|---|---|
Detector | 4 photodiodes + filter | |
Measurement principle | Absorption (Attenuation) | |
Optical path | 0.3 mm, 1 mm, 2 mm, 5 mm, 10 mm, 20 mm, 50 mm | |
Parameters | NO3 / NO3-N, NOx / NOx-N(calibrated with NO3 standard solution) | |
Standard applied | DIN 38404-3, based on DIN 7027 | |
Response time | 40 s |
|---|---|
Temperature compensation | No |
Turbidity compensation | Yes |
Data logger | ~ 2 GB |
Response time T100 | 40 s |
Measurement interval | ≥ 20 s |
Interface | digital | Ethernet (TCP/IP) |
|---|---|---|
RS-485 (Modbus RTU) | ||
Power supply | 12 - 24 VDC (± 10 %) | |
Power consumption | ≤ 7 W | |
Connection | Fixed cable with 8pin M12 connection | ||
|---|---|---|---|
Housing material | Stainless steel (1.4571/1.4404) or titanium (3.7035) | ||
Dimensions (L x Ø) | ~ 470 mm x 48 mm* | ~ 18.5" x 1.9" * | |
Weight | VA | ~ 3 kg | ~ 6.6 lbs |
Ti | ~ 2 kg | ~ 4.4 lbs | |
Sample temperature | 0...+55 °C* | ~ +32 to +131 °F* | |
|---|---|---|---|
Ambient temperature | 0...+55 °C* | ~ +32 to +131 °F* | |
Relative humidity | 0 ... 95 % non-condensing | ||
Storage temperature | -20...+80 °C | ~ -4 to +176°F | |
Max. Pressure | with SubConn | 30 bar | ~ 435 psi |
|---|---|---|---|
with fixed cable | 3 bar | ~ 43.5 psi | |
in FlowCell | 1 bar, 2...4 L/min | ~ 14.5 psi at 0.5 to 1.0 gpm | |
Inflow velocity | 0.1...10 m/s | ~ 0.33 t 33 fps | |
Degree of protection | Sensor side | IP68 | |
Controller side | IP65/ IP67 | ||
Operating altitude | max. altitude 2000 m (6562 ft) | ||
Maintenance effort | ≤ 0.5 h/month typical |
|---|---|
Calibration/ | 24 months |
System compatibility | TriBox3, TriBox mini, Modbus RTU |
Warranty | 1 year (EU & US: 2 years) |