Technical specifications

Measurement technology

Change in inductance

Measurement principle

Change in inductance with two toroidal coils

Parameters

Conductivity

Measurement range

0 – 2000 μS⁄cm

Measuring accuracy

± (2 % + 20 μS/cm)

Drift

0.1 % / year

Temperature compensation

Via NTC

Turbidity compensation

No

Data logger

No

Response time

T90, depending on the equilibrium position

Measurement interval

10 s

Housing material

Noryl

Dimensions (L x Ø)

119 mm x 52 mm

~ 4.6'' x 2.0''

Weight

0.1 kg

~ 0.22 lbs

Interface

RS-485 Modbus RTU (baud rate = 9600)

Power consumption

< 75 mW

Power supply

7 - 40 VDC

Connection

8-pin M12 plug

Maintenance effort

≤ 0.5 h/month typical

Maintenance interval

24 months

Calibration method

Two-point calibration in air and with standard measuring solution during initial installation, followed by validation

System compatibility

Modbus RTU

Warranty

1 year (EU & USA: 2 years)

Max. Pressure

with fixed cable

10 bar

150 psi

Degree of protection

IP68

Temperature

Probe

-10 ... +70 °C (max. 85 °C)

14 ... 158 °F (max. 185 °F)

Ambient temperature

-10 ... +70 °C (max. 85 °C)

14 ... 158 °F (max. 185 °F)

Bearing

-20 ... +80 °C

-4 ... 176 °F

Inflow velocity

Max. 3 m/s, uniform and constant flow