Technical specifications
Measurement technology | Membrane-covered, amperometric 2-electrode system |
|---|---|
Measuring principle | Amperometry |
Parameters | Chlorine dioxide |
Measuring range | 0...2 mg/L, 0...20 mg/L |
Accuracy | Measuring range 2 mg/L: | at 0.4 mg/L < 1 % | at 1.6 mg/L < 1 % |
|---|---|---|---|
Measuring range 20 mg/L: | at 1.5 mg/L < 0.1 % | ||
Response time | T90: approx. 1 min | ||
Run-in time | For initial commissioning approx. 2 h | ||
Drift | approx. -1 % per month | ||
Temperature compensation | Automatic, through integrated temperature sensor; temperature jumps must be avoided | ||
Housing material | Microporous hydrophilic membrane, PVC-U, stainless steel 1.4571 | |
|---|---|---|
Dimensions (Lx Ø) | ~ 205 mm x 25 mm | ~ 8.1" x 1" |
Interface | RS-485, Modbus RTU |
|---|---|
Power supply | 9 - 30 VDC, max. 56mA |
Connection | 8pin M12 plug |
Maintenance interval | Check measuring signal: typically once a week Membrane cap change & electrolyte change: depending on the application |
|---|---|
System compatibility | Modbus RTU |
Warranty | 1 year (EU&US: 2 years) on electronics; |
Process pressure: | 1 bar* | ~14.5 psi* |
|---|---|---|
Calibration method | Chlorine determination with DPD-1 method | |
*No pressure surges and/or vibrations
Process temperature | 0...+50 °C* | ~ +32...+113 °F* |
|---|---|---|
Flow rate | Approx. 15...30L/h in FLC-3, slight flow dependency is present | |
pH range | pH 1 ... pH 12, reduced pH value dependency | |
Conductivity | 10 μS/cm...50 mS/cm (seawater) | |
Cross influences | Cl2 does not interfere; O3: factor 25 | |
*no ice crystals in the sample water