enviroFlu
Fluorometer for measurements of PAH (polycyclic aromatic hydrocarbons). Detects oil in water through UV fluorescence.
PAH, oil-in-water using UV fluorescence
enviroFlu HC is the new generation of immersion sensors for measurement of oil-in-water. The used measuring principle of UV fluorescence is much more sensitive than the conventionally used infrared scattering or absorption method. This allows to determine even the slightest traces of PAH’s, for example in drinking water and cooling water condensates.
Application areas include the petrochemical industry, leakage detection in cooling and wastewater streams as well as environmental monitoring. The devices enable both stationary use in shafts, flows or piping, and mobile use through an optional hand-held measuring instrument. An innovative coating reduces fouling of the optical measuring window and minimizes the maintenance.
Areas of application
- without sampling and sample preparation
- without delay
- without reagents
- optical window with nanocoating
- high sensitivity and selectivity
Parameters
- Drinking water
- Waste water
- Airports
- Cooling water
- Flue gas scrubbing
- Refineries
- Pipeline monitoring
- Desalination plants
Technical Data
Light Source | Xenon flash lamp + filter (254 nm) | |
Detector | Photo diode + filter (360 nm) | |
Measurement principle | Fluorescence | |
Parameter | PAH, oil in water | |
Measuring range | enviroFlu HC 500 | |
PAH: 0…50 ppb, 0…500 ppb | ||
Oil in water: 0…1.5 ppm, 0…15 ppm typical | ||
enviroFlu HC 5000 | ||
PAH: 0…500 ppb, 0…5000 ppb | ||
Oil in water: 0…15 ppm, 0…150 ppm typical | ||
Measurement accuracy | enviroFlu HC 500 :0.3 ppb | |
enviroFlu HC 5000: 0.5 ppb | ||
Turbidity compensation | No | |
Data logger | No | |
T100 response time | ≤10 s | |
Measurement interval | ≥5 s | |
Material | Housing: Stainless steel (1.4571/1.4404) or titanium (3.7035) | |
Measuring Head: black POM or PPS with synthetic quartz glass | ||
Dimensions (L x Ø) | 311 mm x 68 mm | ~12.2˝ x 2.6˝ |
Weight stainless steel | ~ 2.7 kg | ~ 6 lbs |
Weight titanium | ~ 1.9 kg | ~ 4.2 lbs |
Interface digital | RS-232 (TriOS) | |
Interface analog | 4…20 mA, 0…5 V | |
Power consumption | ≤3.5 W | |
Power supply | 12…24 VDC (± 10 %) | |
Maintenance effort | ≤0.5 h/month (typical) | |
Calibration/maintenance interval | 24 months | |
System compatibility | Analog Out (0…5 VDC, 4…20 mA) | |
Warranty | 1 year (EU: 2 years) | US: 2 years |
Max. pressure with SubConn | 30 bar | ~ 435 psig |
Max. pressure with fixed cable | 3 bar | ~ 43.5 psig |
Max. pressure in FlowCell | 1 bar, 2…4 L/min | ~ 14.5 psig, 0.5 to 1 gpm |
Deepsea version | 600 bar | ~ 8702,26 psig |
Protection type | IP68 | NEMA 6P |
Sample temperature | +2…+40 °C | ~ +36 °F to +104 °F |
Ambient temperature | -5…+55 °C | ~ +23 °F to +131 °F |
(0…+40 °C for specified accuracy) | (~ 32 °F to 104 °F for specified accuracy) | |
Storage temperature | -20…+80 °C | ~ -4 °F to +176 °F |
Inflow velocity | 0.1…10 m/s | ~ 0.33 fps to 33 fps |
Max. immersion depth | 300 m with SubConn 8-pin underwater connector | ~ 984 ft with SubConn 8-pin underwater connector |
30 m with fixed cable | ~ 98 ft with fixed cable | |
optional: 6000 m Deepsea version | optional:~ 19685,04 ft Deepsea version |